Transferring a 510(k): What FDA’s Final Guidance Requires

Table of Contents

FDA finalized its 510(k) transfer guidance in November 2026. Here is what buyers and sellers of cleared devices must do on listing, numbers and GUDID.

Last updated: November 20, 2026

A 510(k) clearance can be sold or transferred without a new submission if the device is unchanged. FDA’s final guidance says the new holder must register, create a new device listing using the original 510(k) number, and update GUDID, while the previous holder updates or discontinues its listing.

Scope: what FDA’s final guidance “Transfer of a Premarket Notification (510(k)) Clearance, Questions and Answers” requires of the previous and new holder, what changed from the June 2025 draft, and a diligence checklist for device deals.

What did FDA finalize on 510(k) transfers?

FDA’s Center for Devices and Radiological Health finalized a question and answer guidance on how a 510(k) clearance moves from one company to another. The guidance page is dated November 19, 2026, the PDF says it was issued November 20, 2026, and it sits under docket FDA-2024-D-1528. The draft was issued June 5, 2025.

The scope is plain: the guidance “applies to both the previous and the new holder of a 510(k) after the 510(k) clearance for a specific device is sold or transferred from one person to another.” CBER is listed as a contributor, so it also touches devices regulated by the biologics center.

Nothing here creates a new pathway. The guidance explains how existing registration, listing and unique device identification rules in 21 CFR Parts 807 and 830 apply when a clearance changes hands.

Can you sell or transfer a 510(k) clearance?

Yes. A 510(k) clearance can be sold or transferred, and if the device is not modified, the new holder does not need to submit a new 510(k), according to the guidance, which cites 21 CFR 807.81(a).

That is why cleared devices trade as assets in medtech M&A, carve outs and product line sales. The value of a clearance sits in the regulatory work already done. The guidance confirms that value survives a transfer, as long as the buyer keeps the device as cleared and does its own paperwork.

Who is the 510(k) holder after a transfer?

Only one company at a time. FDA’s answer is “No” to the question of whether more than one company can hold a 510(k): FDA “considers there to be a single ‘510(k) holder'” for a device at a time.

Two details trip people up. First, the applicant named on the original clearance does not change in FDA’s 510(k) database even after a transfer; a footnote says the person listed as the applicant “does not change even if the 510(k) is transferred.” Second, FDA identifies the current holder through device listing, not through the database entry. So the listing is the record that matters.

What does the new 510(k) holder have to do?

The new holder registers its establishment, creates a new device listing that cites the original 510(k) number, and updates labeling and GUDID where the labeler information changed. Each step has a regulatory hook.

Step Previous holder New holder Rule cited in guidance
Registration Keeps or updates its own registration Registers within 30 days after entering the operation 21 CFR 807.22(a)
Device listing Updates listing if it stops an activity; discontinues if it stops distribution Creates a new listing; system assigns a new listing number 21 CFR 807.25, 807.28(a), 807.28(d)
Premarket number None Enters the original 510(k) number if the device is not significantly changed 21 CFR 807.25(g)(4)
Annual update Through its own listing Every fiscal year, Oct 1 to Dec 31 21 CFR 807.22(b)(3)
Labeling None Label updates may be required 21 CFR 801.1(a)
GUDID None Update labeler data; by first labeling with the change, or within 10 business days if not on the label 21 CFR 830.310, 830.330(b)

The guidance is explicit that this is the company’s job. Updating registration or listing is “the responsibility of the owner or operator of the establishment,” and FDA is “not required to update registration or listing information in response to such notifications.” Sending FDA a letter about the deal does not do the work.

When does a transferred device need a new 510(k)?

When the device is significantly changed. If the new holder modifies the device in a way that triggers 21 CFR 807.81(a)(3), it “must generally submit a new 510(k)” and use the new number in its listing after clearance.

The guidance also restates the timing rule for new submissions: generally submit “at least 90 days before proposing to begin the introduction or delivery for introduction into interstate commerce.” For buyers planning to rebrand, change materials or update software right after closing, that clock belongs in the integration plan. Document the analysis for every planned change.

Who else uses the original 510(k) number?

Contract manufacturers, contract sterilizers, repackagers and relabelers are required to use the FDA assigned premarket submission number of the cleared 510(k) in their listings, citing 21 CFR 807.25(g)(4). A transfer therefore ripples through the supply chain: every partner listing the device should reference the right clearance after the deal.

Initial importers have a different option. They may list the manufacturer’s name and address instead, under 21 CFR 807.20(a)(5). Foreign establishments are covered too; the guidance notes a manufacturer “can be either domestic or foreign.”

What changed from the June 2025 draft?

The final is shorter and reorganized. Regulatory Focus (RAPS) reported on November 19, 2026 that the final “significantly shortens certain sections, while adding new questions and answers,” with new questions apparently drawn from the draft’s background section.

RAPS also reported that the final “makes no distinction between domestic and foreign manufacturers” and clarifies when a manufacturer must submit a 510(k). Those comparisons are attributed to RAPS; we read the final text, not a line by line redline.

What happens if you skip the transfer steps?

The device may be considered adulterated or misbranded. The guidance says: “If you fail to register, list, update GUDID information, or submit a marketing application when required, your device may be considered adulterated and/or misbranded,” citing FD&C Act sections 501(f)(1)(B), 502(a)(1), 502(o) and 502(t)(2).

Our view: the first risk is often commercial. Customers and distributors can check listings and UDI data. A device whose listing still points at the seller, or whose GUDID record names the wrong labeler, invites questions during contracting and recalls. Recalls also depend on clean records of who is responsible for the device.

How do you get the 510(k) records after buying a device?

Ask the seller, or file a FOIA request. The guidance says the new holder can obtain records from the prior owner or through FOIA, and notes “CDRH posts 510(k) records previously released to the public through FOIA requests in the 510(k) database.”

Our view: never close a device deal without the full 510(k) file, design history file and complaint history in hand. FOIA copies are redacted and slow. The asset purchase agreement should require delivery of the complete submission and correspondence at closing.

510(k) transfer checklist for device deals (our framework)

This is our working checklist for regulatory, quality and business development teams on either side of a device transaction. It is built from the guidance above, not a substitute for counsel.

  • Confirm what is being transferred. List every 510(k) number, the device as cleared and any modifications made since clearance, with letters to file.
  • Get the full file. Original submission, FDA correspondence, design history file and complaint and MDR history at closing.
  • Register within 30 days. The buyer registers its establishment within 30 days after entering the operation.
  • List with the original number. Create the new listing citing the original 510(k) number; note the new listing number.
  • Make the seller update or discontinue. Put the seller’s listing update or discontinuation into the closing deliverables.
  • Update GUDID and labels. Change labeler data on the timeline in 21 CFR 830.330(b) and plan label changes.
  • Notify supply chain partners. Contract manufacturers, sterilizers, repackagers and relabelers must reference the right clearance.
  • Screen planned changes. Any post close change that is significant needs a new 510(k), generally 90 days before marketing.
  • Calendar the annual update. Oct 1 to Dec 31 each fiscal year.

What to watch next

  • Whether FDA’s registration and listing system adds tools to show the current holder more clearly in public records.
  • How medtech carve outs and portfolio sales use the guidance in transition service agreements.
  • Any related CDRH guidance on device modifications and when a change needs a new 510(k).

FAQ

Can a 510(k) be transferred to another company?

Yes. FDA’s final guidance confirms a 510(k) clearance can be sold or transferred, and the new holder does not need a new 510(k) if the device is unchanged.

Do I need to notify FDA when I buy a 510(k)?

The guidance puts the work on the companies. The new holder must register and list the device itself, and FDA says it is not required to update registration or listing information in response to notifications.

How long does the new owner have to register?

The guidance says to register within 30 days after entering into the operation and to submit listing information including the 510(k) number.

Does the 510(k) number change after a transfer?

No. If the device is not significantly changed, the new holder enters the original FDA assigned 510(k) number. The listing number is new.

Does FDA’s 510(k) database show the new owner?

No. The applicant on the original clearance does not change in the database. FDA identifies the current holder through device listing.

Can two companies hold the same 510(k)?

No. FDA considers there to be a single 510(k) holder for a device at a time.

What must the seller do after a 510(k) transfer?

Update its device listing if it stops an activity, and discontinue the listing if it stops distributing the device, under 21 CFR 807.28.

Do I have to update GUDID after buying a device?

Yes, if labeler information such as the labeler name changed. Update by the date a device is first labeled with the change, or within 10 business days if the information is not on the label.

When is a new 510(k) required after a transfer?

When the new holder significantly changes the device under 21 CFR 807.81(a)(3). Submissions generally go in at least 90 days before marketing.

What happens if a buyer does not register or list?

FDA says the device may be considered adulterated or misbranded under the FD&C Act.

How do I get a copy of a 510(k) I acquired?

From the prior owner, or through a FOIA request. CDRH posts previously released 510(k) records in its database.

When was the 510(k) transfer guidance finalized?

The guidance page is dated November 19, 2026 and the PDF says it was issued November 20, 2026. The draft came out June 5, 2025.

Sources

  • FDA, “Transfer of a Premarket Notification (510(k)) Clearance, Questions and Answers,” final guidance, November 2026, docket FDA-2024-D-1528 (primary): guidance page and PDF
  • Regulatory Focus (RAPS), Ferdous Al-Faruque, Nov 19, 2026 (attributed: changes from the draft): raps.org